Drop test before or after the IP test: why the order decides

Bench with a drop rig at one end and an immersion tank at the other, sample device between them

Quick summary

  • Testing IP and drop on the same sample means the order decides the result: drop first, then IP, is a much harder test, because a drop deforms exactly the features that keep water out.
  • The other direction has its own problem: water left inside after IP testing can distort a drop test, or cause a short circuit when power is applied.
  • IEC 60529 does not define the order. The order belongs to the test programme or the contract — so if it is not written, each side reads it differently.
  • The safest approach: state the order in the test request, state the recovery time between tests, and use two separate samples when a firm conclusion matters.

A common request at the lab: “One sample, please run drop and IP together.” The right answer is not yes or no, but: in which order. This article explains the mechanics and how to write it clearly.

1. The mechanics: a drop damages exactly what seals

A drop does more than scratch the housing. It acts on the four things an IP test cares about most:

Feature affected Mechanism Effect on the IP test
Gasket / seal Deformed, twisted, popped out of its groove A new gap — water and dust get in more easily
Plastic or metal housing Micro-cracks, deformed corners, misaligned flanges Leak paths invisible to the eye
Screws, clips, snaps Cracked screw bosses, released snaps Lower gasket compression, wider gap
Speaker mesh, membrane vent Torn, stretched, adhesive lifted Loses water blocking but may still block dust (or the reverse)
Close-up of a device housing corner after a drop, small crack in the plastic near a screw boss, scuffed edge
A crack around a screw boss is the classic leak path: hard to see, easy for water to find.

So drop first, then IP is the harder test — and the more realistic one: equipment is dropped before it meets rain, not after.

2. Three orders and what each one means

Order Question answered Severity Used when
IP first → drop after The housing seals; and once sealed, does it survive a drop Lower for the IP claim Confirming basic capability of the design
Drop first → IP after After impact, does the housing still seal Higher, closer to field reality Handheld and field equipment
Repeated cycles After many drops and many wettings, does it still seal Highest Long-term outdoor products

The key point: “IP67 passed” under the first two orders cannot be compared directly. Both reports say IP67, but one is an intact housing and the other has been through impact. If a buyer compares two suppliers without checking the order, the comparison is meaningless.

3. The other direction: water left inside

Testing IP first and dropping afterwards brings its own problems — and this is where many labs skip a step:

  • Residual water in gaps, in the speaker mesh, in the battery compartment changes mass and load distribution during the drop.
  • Corrosion starts immediately after water testing; waiting a few days before the drop means testing a different sample.
  • Powering up after immersion for a functional check can cause a short — dry and recover for the agreed period first.
  • Water can freeze if the next test is low temperature, distorting the result entirely.
Close-up of a rubber gasket compressed in a housing groove, slightly twisted out of position, dust specks visible
A gasket twisted in its groove may be an assembly error — or the aftermath of an earlier drop.

4. What the standards say about order

An important and often misunderstood point: IEC 60529 does not define the order between the IP test and mechanical or environmental tests. The standard describes how to run the IP test on a sample; it does not describe a product’s test sequence.

The order comes from three sources:

Source Role
Product or sector standard May prescribe a specific sequence for that product type
Customer test programme (tailoring) Decides order, number of cycles, recovery time
Contract / test request If the two above are silent, the contract must speak

In programmes following MIL-STD-810, the sequence is part of tailoring and must appear in the test plan. For consumer products, in most cases nobody specifies it — which is exactly when the test request has to.

5. How to write it in the test request

What to state Example
Order “Drop test first, then IP on the same sample”
Number of samples per test “01 sample for drop → IP; 01 control sample for IP only”
Recovery time “Recover 24 h at room conditions between the two tests”
Post-water handling “Wipe the surface dry, no heat drying, do not dismantle the sample”
Acceptance criteria “Pass if no water in the electronics compartment; record water in the battery compartment either way”
Report content “The report must state the order and the recovery time”

Model wording: “Testing on the same sample in the following order: (1) drop per …; (2) recover 24 h at 23 ± 5 °C; (3) IPX7 test per IEC 60529. The report shall state the order, the recovery time and the sample condition before each test.”

6. Why use two samples

If the goal is a firm conclusion rather than saving samples, use two separate samples for the two tests plus a third for the combined sequence:

Sample Used for Conclusion obtained
Sample A IP only Sealing capability of the intact design
Sample B Drop only Impact capability, undistorted by water
Sample C Sequence drop → IP Whether the design still seals after impact

Three samples give three different answers. One sample gives one answer — usually the hardest one to interpret when something goes wrong.

Device drying on a rack after water testing, droplets on the housing, small fan airflow, technician placing a second sample
Recovery time between tests belongs in the test request, not in the lab’s discretion.

7. Designing a repeated cycle

For long-life outdoor products, one drop and one wetting are not enough. A repeated cycle puts the two tests next to each other many times, and the interesting figure is not “did it pass once” but how many rounds it survived.

Step in one cycle Purpose Must be stated
Drop on the selected faces Creates cumulative deformation Faces, height, impact surface
Visual inspection Detects cracks, released snaps, shifted gaskets Photos from every round
IP test at the chosen level Checks whether it still seals Level, conditions, duration
Recovery and functional check Separates mechanical from electrical faults Recovery time, pass criteria

The number of rounds, the drop height and the IP level in each round have no universal values — they must come from the real service environment and be written down before testing. One thing should be fixed in advance: stop at first failure, or run all rounds and then assess. The two choices yield different kinds of conclusion.

8. Six common mistakes

Mistake Consequence
No order stated, each side assumes one Dispute when the result is not as expected
IP tested on a dropped sample but reported as intact Result not reproducible, certificate loses value
Drop immediately after immersion with no recovery Water inside distorts the result
Reusing an IP-tested sample without recording it Nobody knows what the sample has been through
No record of sample condition before each test Cannot separate design fault from cumulative damage
Comparing two suppliers with different test orders Wrong conclusion about relative quality
Running cycles without a pre-agreed stopping rule Cannot state “failed at round N”

Frequently asked questions

Does IEC 60529 require drop before or after the IP test?

Neither. IEC 60529 only describes how to run the IP test on a sample. The order belongs to the product standard, the test programme or the contract.

Does drop-first-then-IP make the result “worse”?

Yes, and that is the point: it reproduces the real sequence. But it must be stated in the report, otherwise it will be read as a result on an intact sample.

Can one sample do both tests?

Technically yes, but it gives one answer and is hard to interpret when something fails. Add a control sample.

How long should recovery after water testing be?

The programme decides, and it must be written down. If nothing specifies it, agree it in writing before testing — do not leave it to the lab.

Should the device be functionally checked after each test?

Yes, with stated criteria. Functional checks after water testing should happen after recovery, to avoid short circuits.

Conclusion

Test order is not an administrative detail. It changes the result, changes how the result is interpreted, and changes the legal value of the report.

Three things to do: state the order in every test request that includes more than one test; state the recovery time between tests; and for important products, use control samples to separate “design fault” from “cumulative damage”.

References

  • IEC 60529 — degrees of protection (IP code); scope: testing on a sample, not a product test sequence.
  • IEC 60068-2-27 — shock test within the environmental testing series.
  • MIL-STD-810 — tailoring principles and test sequence within a programme.

Related articles


Talk to us


    Disclaimer

    This article is our own interpretation; it is not legal advice. Readers should check the official text of the standard before applying it to a specific product.